• <tr id='MLWCVp'><strong id='MLWCVp'></strong><small id='MLWCVp'></small><button id='MLWCVp'></button><li id='MLWCVp'><noscript id='MLWCVp'><big id='MLWCVp'></big><dt id='MLWCVp'></dt></noscript></li></tr><ol id='MLWCVp'><option id='MLWCVp'><table id='MLWCVp'><blockquote id='MLWCVp'><tbody id='MLWCVp'></tbody></blockquote></table></option></ol><u id='MLWCVp'></u><kbd id='MLWCVp'><kbd id='MLWCVp'></kbd></kbd>

    <code id='MLWCVp'><strong id='MLWCVp'></strong></code>

    <fieldset id='MLWCVp'></fieldset>
          <span id='MLWCVp'></span>

              <ins id='MLWCVp'></ins>
              <acronym id='MLWCVp'><em id='MLWCVp'></em><td id='MLWCVp'><div id='MLWCVp'></div></td></acronym><address id='MLWCVp'><big id='MLWCVp'><big id='MLWCVp'></big><legend id='MLWCVp'></legend></big></address>

              <i id='MLWCVp'><div id='MLWCVp'><ins id='MLWCVp'></ins></div></i>
              <i id='MLWCVp'></i>
            1. <dl id='MLWCVp'></dl>
              1. <blockquote id='MLWCVp'><q id='MLWCVp'><noscript id='MLWCVp'></noscript><dt id='MLWCVp'></dt></q></blockquote><noframes id='MLWCVp'><i id='MLWCVp'></i>

                【高低温试验箱】高低温恒温试验箱,高低温环境试验箱,高低温测试仪器-林频仪器股份有限公司

                林频·打造行业高品质试验设备 用工匠精神铸就环境试验箱
                助力您的产品腾飞

                服务热线:4000-662-888

                您的位置: 上海林频仪器股份有限公司 > 新闻中心 > 常见问题 > 恒温恒湿试验试验箱湿度显示100%或做低湿点时,湿度下不来?

                恒温恒湿试验试验箱湿度显示100%或做低湿点时,湿度下不来?

                返回列表 来源:林频合作伙伴 发布日期:2019-06-05 16:59

                  按以下步骤逐一排查:

                  (A)湿球纱布发黄,或者使用时间过久,纱布上积淀太多矿物质,导致湿球纱布吸水性能下降,吸不上水,从而使检测到的湿度不准确,造成湿度偏高的假像;

                  (B)纱布挂错位置,工作室内有两支传感器,上面一支传感器为干球,下面一支传感器为湿球,正常情况下,必须将纱布挂在下面的一支传感器上,且纱布下端必须可靠的浸在积水头水槽内,以便湿球纱布能可靠的吸水。